用光腔衰荡技术测量镜片的反射率
Analysis of reflectivity measurement by cavity ring-down spectroscopy
-
摘要: 根据光腔衰荡光谱技术原理,建立了测量镜片反射率的实验装置。利用该装置测定了一对反射率相等的高反射腔镜,反射率测试结果为(99.925±0.001)%;以22.5°将直腔变为折叠腔,测得的反射率为(99.992±0.003)%。重复测定反射镜样品的反射率,精度达到10-5。该测量装置可用于超低损耗薄膜高反射镜反射率的精确测量。轻微移动探测器位置,对腔镜的测试结果影响不大。
-
关键词:
- 光腔衰荡光谱技术(CRDS) /
- 反射率 /
- 衰荡时间 /
- 直腔 /
- 折叠腔
Abstract: This paper presents the experimental method to measure the high reflectivity of optical mirror by cavity ring-down spectroscopy. The experimental system uses a straight cavity composed of a pair of cavity mirrors with same reflectivity or a folded cavity which is formed from the straight cavity to be folded at 22.5° angle, a reflector is laid at folded point.The reflectivity of the cavity mirrors is measured to be(99.925±0.001)% and the reflector to be (99.992±0.003)%.The metrical precision achieves 10-5.The low loss and high reflectivity mirrors can be measured accurately by this setup.-
Key words:
- cavity ring-down spectroscope (crds) /
- reflectivity /
- ring-down time /
- straight cavity /
- folded cavity
点击查看大图
计量
- 文章访问数: 2439
- HTML全文浏览量: 337
- PDF下载量: 618
- 被引次数: 0