自发喇曼成像法在线测量流场组分浓度(II)
Measurements of the species concentration in flow field with the technique of spontaneous Raman imaging (II)
-
摘要: 介绍了利用自发Raman散射光谱测量低压气流组分绝对浓度的方法,在流动状态下得到了不同压力下的自发喇曼散射光谱,检测最低压力为133帕;对于含N2氧气流在已知N2浓度时实时标定检测了O2的绝对浓度,测量误差小于8%。
-
关键词:
- Raman散射绝对浓度单重态氧发生器
Abstract: A spontaneous Raman imaging system which can monitor the concentrations of major species in flow field in real time has been described. The results of the low pressure O2 are presented and the lowest pressure we measure is 133Pa. This technique can be calibrated in real time by an introduction of a known amount of N2. The error is less than 8%. The measurement precision of O2(a1Δ) concentration will be improved largely when this technique is applied to monitor chemical oxygen-iodine laser singlet oxygen generator performance.-
Key words:
- spontaneous raman scattering
点击查看大图
计量
- 文章访问数: 1775
- HTML全文浏览量: 303
- PDF下载量: 338
- 被引次数: 0