多通道软X射线Dante谱仪标定及实验
Calibration and application of multi-channel soft X-ray Dante spectrometer
-
摘要: 介绍了用于阳加速器上Z箍缩内爆实验诊断的Dante谱仪的结构和通道配置,详细讨论了X射线二极管、掠入射平面反射镜和滤片等主要元器件的标定结果,分析了标定结果与理论计算发生偏差的原因,给出了喷气Z箍缩等离子体辐射的测量结果,X光辐射功率30~40 GW,能量约0.8 kJ,并与闪烁体光电管测量结果进行了比较,差异约20%。Abstract: In Z-pinch diagnostic, Dante spectrometer plays an important role in the measurement of soft X-ray spectrum and total flux. This paper reports on the structure and channel configuration of a “Dante” spectrometer designed for z-pinch experiments on Yang accelerator. Calibration of X-ray devices used in the spectrometer, such as X-ray diodes, grazing incidence mirrors and filters are described in detail and reasons accounting for the deviation from theoretical values have also been discussed. Waveforms and unfolding results which also compared with X-ray scintillator powermeter obtained in an argon gas-puff experiment are presented as well, the power of X- ray radiation is about 30~40 GW and the energy of it is about 0.8 kJ, the difference of the results of Dante spectrometer and X-ray sc
-
Key words:
- z-pinch /
- soft x-ray /
- dante spectrometer /
- calibration