基于252Cf中子源自相关分析的核事件死时间测量与修正方法
Novel method for measuring and correcting dead-time using 252Cf neutron source based autocorrelation function in nuclear radiation detection
-
摘要: 基于所研制的252Cf中子源频域测量系统,针对核探测系统的死时间使核事件丢失这一问题,利用自相关分析,研究了一种核探测系统死时间测量与修正的方法。试验测量结果表明:采用自相关图谱,可以直接获得高速频谱测量系统的死时间为40 ns,既方便又快捷,且测量精度高,达到了ns级,优于国外文献报道的类似核武器识别测量系统的死时间20 μs。Abstract: To solve the problem of dead time loss, a novel method for measuring and correcting dead-time is proposed with the use of autocorrelation function based on a 252Cf -source-driven spectrum analyzing system. The experiment shows that the dead-time of the analyzing system is 40 ns. The dead-time can be obtained directly from the spectrum of autocorrelation with high measuring precision of nanosecond level.
-
Key words:
- nuclear radiation detection /
- dead-time /
- autocorrelation /
- correction /
- neutron source
点击查看大图
计量
- 文章访问数: 1891
- HTML全文浏览量: 178
- PDF下载量: 432
- 被引次数: 0