杆箍缩二极管X射线焦斑的成像法测量
Imaging measurement of X-ray spot of rod-pinch diode radiographic source
-
摘要: X射线源的焦斑尺寸是反映杆箍缩二极管射线源成像性能的重要参数。利用针孔成像法对MeV级脉冲X射线源的焦斑进行了2维图像测量。厚针孔采用直孔段加单锥体结构,直孔段孔径为0.2 mm。对于0.5 MeV的X射线,5倍成像倍率下调制传递函数值为0.5时空间分辨达到2.0 lp·mm-1。图像采集系统由闪烁体、物镜和CCD相机组成,物镜的成像倍率约0.34。实验结果经过模糊校正后,得到了焦斑的图像和调制传递函数。根据调制传递函数值为0.5时对应的空间频率值,给出X射线源焦斑的尺寸。阳极杆直径为1.2 mm时,X射线源焦斑的高斯分布等效直径为0.86 mm。Abstract: A pulsed X-ray radiographic source based on inductive voltage adder (IVA) and rod-pinch diode (RPD) is developed at the Northwest Institute of Nuclear Technology. The X-ray spot size of RPD is a critical parameter to evaluate the spatial resolution performance for radiography. Using thick pinhole imaging technique, two-dimension images of the RPD X-ray spots have been measured. A pinhole with 0.2-mm-diameter was designed and an image acquirement system based on a lutetium-oxyorthosilicate-crystal scintillator, a lens and a CCD camera was established. The magnifications of the pinhole and the lens are about 5 and 0.34, respectively. The spatial resolution of the pinhole imaging layout is 2.0 lp·mm-1 at the modulation transfer function (MTF) of 0.5 for 0.5 MeV X-ray. Through the restoration
-
Key words:
- rod-pinch diode /
- pulsed x-ray source /
- spot size /
- pinhole /
- imaging measurement