Design principle and performance of RIXS spectrometer

doi: 10.11884/HPLPB201830.170153
  • Received Date: 2017-12-05
  • Rev Recd Date: 2018-03-22
  • Publish Date: 2018-07-15
  • RIXS, Resonant Inelastic X-ray Scattering, is an x-ray spectroscopy technique used to investigate the electronic structure of molecules and materials. The requirement of it for source and spectrometer is extremely strict. Mastering its design idea, application method, performance, and so on is of great significance for future research. Through analyzing the design ideas of RIXS spectrometer, using LPF of theory analysis and calculating the varied-line-spacing grating parameters, this paper analyzes each aberration effect on the result of the energy resolution. The relationship between motor encoder and accurate angle of each part of spectrometer is needed during setting up and operating. Thus a non-contact measurement using visible laser was applied and it has been proved to be effective. Before the first commissioning of XFEL, the spectrometer was calibrated by X-ray in synchrotron end station for its energy resolution. The SHADOW program simulation shows that the performance of spectrometer meets the requirement of design and experiment.
  • loading
  • [1]
    Vankó G, Bordage A, Glatzel P, et al. Spin-state studies with XES and RIXS: From static to ultrafast[J]. Journal of Electron Spectroscopy and Related Phenomena, 2013, 188: 166-171. doi: 10.1016/j.elspec.2012.09.012
    [2]
    丁大军, 钱莉民. 软X射线共振非弹性光散射及其应用[J]. 光散射学报, 2000, 12(4): 211-215. doi: 10.3969/j.issn.1004-5929.2000.04.003

    Ding Dajun, Qian Limin. Soft X-Ray resonant inelastic scattering and its applications. The Journal of Light Scattering, 2000, 12(4): 211-215 doi: 10.3969/j.issn.1004-5929.2000.04.003
    [3]
    Strocov V N, Schmitt T, Flechsig U, et al. High-resolution soft X-ray beamline ADRESS at the Swiss Light Source for resonant inelastic X-ray scattering and angle-resolved photoelectron spectroscopies[J]. Journal of Synchrotron Radiation, 2010, 17(5): 631-643. doi: 10.1107/S0909049510019862
    [4]
    McKinney W R. Varied line-space gratings and applications[J]. Review of Scientific Instruments, 1992, 63(1): 1410-1414. doi: 10.1063/1.1143030
    [5]
    吴军, 杨家敏, 丁永坤, 等. 平焦场光栅光谱仪[J]. 强激光与粒子束, 2002, 14(4): 489-492. http://www.hplpb.com.cn/article/id/1345

    Wu Jun, Yang Jiamin, Ding Yongkun, et al. Flat field grating spectrograph for soft X-ray laser research. High Power Laser and Particle Beams, 2002, 14(4): 489-492 http://www.hplpb.com.cn/article/id/1345
    [6]
    Namioka T, Noda H, Seya M. Possibility of using the holographic concave grating in vacuum monochromators[J]. Sci Light, 1973, 22: 77-99.
    [7]
    Noda H, Namioka T, Seya M. Geometric theory of the grating[J]. J Opt Soc Am, 1974, 64(8): 1031-1036. doi: 10.1364/JOSA.64.001031
    [8]
    Miyamoto K. On a comparison between wave optics and geometrical optics by using Fourier analysis. Ⅲ. Image evaluation by spot diagram[J]. J Opt Soc Am, 1959, 49(1): 35-40. doi: 10.1364/JOSA.49.000035
    [9]
    Namioka T. Theory of the concave grating. I[J]. J Opt Soc Am, 1959, 49(5): 446-460. doi: 10.1364/JOSA.49.000446
    [10]
    Harada T, Kita T. Mechanically ruled aberration-corrected concave gratings[J]. Applied Optics, 1980, 19(23): 3987-3993. doi: 10.1364/AO.19.003987
    [11]
    Kita T, Harada T, Nakano N, et al. Mechanically ruled aberration-corrected concave gratings for a flat-field grazing-incidence spectrograph[J]. Applied Optics, 1983, 22(4): 512-513. doi: 10.1364/AO.22.000512
    [12]
    Welnak C, Chen G J, Cerrina F. SHADOW: A synchrotron radiation and X-ray optics simulation tool[J]. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1994, 347(1/3): 344-347.
    [13]
    Sanchez del Rio M, Canestrari N, Jiang F, et al. SHADOW3: A new version of the synchrotron X-ray optics modelling package[J]. Journal of Synchrotron Radiation, 2011, 18(5): 708-716. doi: 10.1107/S0909049511026306
    [14]
    Watanabe M, Toyoshima A, Azuma Y, et al. Performance of a varied-space plane-grating monochromator[C]//Optical Science, Engineering and Instrumentation'97. International Society for Optics and Photonics, 1997: 58-66.
    [15]
    李朝阳. 变线距光栅单色器设计及关键技术[D]. 合肥: 中国科技大学, 2014: 22-23.

    Li Zhaoyang. Design and key techniques of varied line-space grating monochromator. Hefei: University of Science Technology of China, 2014: 22-23
    [16]
    Rubensson J E. RIXS dynamics for beginners[J]. Journal of Electron Spectroscopy and Related Phenomena, 2000, 110: 135-151.
    [17]
    张文海, 曹磊峰, 朱效立, 等. 高线密度X光透射光栅衍射效率[J]. 强激光与粒子束, 2012, 24(10): 2347-2350. doi: 10.3788/HPLPB20122410.2347

    Zhang Wenhai, Cao Leifeng, Zhu Xiaoli, et al. Diffraction efficiency of high line-density X-ray transmission gratings. High Power Laser and Particle Beams, 2012, 24(10): 2347-2350 doi: 10.3788/HPLPB20122410.2347
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Figures(12)  / Tables(2)

    Article views (1885) PDF downloads(420) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return