Yuan Jiandong. Summary of application of stretched wire technology in accelerator alignment and survey[J]. High Power Laser and Particle Beams, 2020, 32: 045102. doi: 10.11884/HPLPB202032.190259
Citation: Yuan Jiandong. Summary of application of stretched wire technology in accelerator alignment and survey[J]. High Power Laser and Particle Beams, 2020, 32: 045102. doi: 10.11884/HPLPB202032.190259

Summary of application of stretched wire technology in accelerator alignment and survey

doi: 10.11884/HPLPB202032.190259
  • Received Date: 2019-07-09
  • Rev Recd Date: 2019-12-17
  • Publish Date: 2020-03-06
  • To study the application and development of stretched wire technology in accelerator alignment and survey, the principle of stretched wire technology is summarized firstly, then the development history of stretched wire alignment method in accelerator is reviewed, and the latest development trend of stretched wire alignment method is introduced. Finally, the advantages and disadvantages of various alignment methods and their suitable environments are discussed, and the development differences between domestic and foreign alignment methods are compared. The development direction of the stretched wire alignment method is pointed out, and the direction of the accelerator alignment and survey is given.
  • loading
  • [1]
    亚姆巴耶夫. 精密基准线测量[M]. 北京: 测绘出版社, 1981: 140-146.

    Ambayev X K. Precision baseline survey[M]. Beijing: Surveying and Mapping Publishing House, 1981: 140-146
    [2]
    鲍利沙科夫. 精密工程测量的方法和仪器[M]. 北京: 测绘出版社, 1981: 202-209.

    Paulishakov V D. Methods and instruments for precision engineering surveying[M]. Beijing: Surveying and Mapping Publishing House, 1981: 202-209
    [3]
    GB/T 15314-1994, 精密工程测量规范[S].

    National Standards of the People's Republic of China. GB/T_15314-1994 Specification for precision engineering surveying[S].
    [4]
    张正禄, 吴栋材, 杨仁. 精密工程测量[M]. 北京: 测绘出版社, 1992: 129-133, 174-177.

    Zhang Zhenglu, Wu Dongcai, Yang Ren. Precision engineering survey[M]. Beijing: Surveying and Mapping Publishing House, 1992: 129-133, 174-177
    [5]
    吴翼麟, 孔祥元. 特种精密工程测量[M]. 北京: 测绘出版社, 1993: 25-26, 210-221.

    Wu Yilin, Kong Xiangyuan. Special precision engineering survey[M]. Beijing: Surveying and Mapping Publishing House, 1993. 11: 25-26, 210-221
    [6]
    华锡生, 黄腾. 精密工程测量技术及应用[M]. 南京: 河海大学出版社, 2008: 57-62, 94-95.

    Hua Xisheng, Huang Teng. Precision engineering survey technology and application[M]. Nanjing: Hehai University Press, 2008: 57-62, 94-95
    [7]
    李广云, 范百兴. 精密工程测量技术及其发展[J]. 测绘学报, 2017, 46(10):1742-1751. (Li Guangyun, Fan Baixing. The development of precise engineering surveying technology[J]. Acta Geodaetica et Cartographica Sinica, 2017, 46(10): 1742-1751 doi: 10.11947/j.AGCS.2017.20170313
    [8]
    赵吉先, 刘荣, 郑加柱, 等. 精密工程测量[M]. 北京: 科学出版社, 2010: 106-112.

    Zhao Jixian, Liu Rong, Zheng Jiazhu, et al. Precision engineering survey[M]. Beijing: Science Press, 2010: 106-112
    [9]
    Fischer G E. Alignment and vibration issues in tev linear collider design[C]//Proceedings of the 1st International Workshops on Accelerator Alignment. 1989: 12-19.
    [10]
    Gordon B. Stretched wire mechanics[C]//Proceedings of the 8th International Workshops on Accelerator Alignment. 2004: 24-27
    [11]
    Touzé T. Effects of the Earth inertial forces on the Wire Positioning System[R]. TS-Note-2007-002.
    [12]
    Goldstine H H. A history of the calculus of variations from the 17th through the 19th Century (Studies in the history of mathematics and physical Sciences[M]. Berlin: Springer, 1980: 35-39.
    [13]
    Behroozi F, Mohazzabi P, McCrickard J P. Remarkable shapes of a catenary under the effect of gravity and surface tension[J]. Am J Phys, 1994, 62(12): 1121-1128. doi: 10.1119/1.17672
    [14]
    Mareno A, English L Q. The stability of the catenary shapes for a hanging cable of unspecified length[J]. European Journal of Physics, 2009(30): 97-108.
    [15]
    Jean P Q, Hélène M D, Thomas T. Stretched wire offset measurements: 40 years of practice of this technique at CERN[C]//Proceedings of the 11th International Workshops on Accelerator Alignment. 2010, 11(2): 23-29.
    [16]
    ЗацаРиннЫй. Automation of precision engineering surveying[M]. Beijing: Surveying and Mapping Publishing House, 1981: 108-119.
    [17]
    GB/T 22543.1-2008, 大坝监测仪器引张线仪[S].

    GB/T 22543.1-2008, Dam monitoring instrument extension line instrument[S]
    [18]
    Fuchs J F, Rude V, Duquenne M, et al. Validation of wire measurements in the LHC tunnel[C]//Proceedings of the 15th International Workshops on Accelerator Alignment. 2018, 15(2): 24-29.
    [19]
    Liska D, Daue1sberg L B, Spalek G, et al. Precision alignment of permanent magnet drift tubes[C]//Proceedings of the 1986 International Linac Conference. 1986, 12(2): 17-28.
    [20]
    Gervaise J, Wilson E J. High precision geodesy applied CERN accelerators[C]//Cern Accelerator School, Applied Geodesy for Particles Accelerators. 1986: 161-162.
    [21]
    Loffler F. Referencing the magnetic axis for HERA’s superconducting magnets[C]//Proceedings of the 1st International Workshops on Accelerator Alignment. 1989, 1(3): 13-19.
    [22]
    潘正风. 北京正负电子对撞机直线加速器安装的精密测量[J]. 武汉测绘科技大学学报, 1991, 16(1):13-18. (Pan Zhengfeng. Precision measurement of the linac installation of Beijing Positron-Negative Collider[J]. Journal of Wuhan University of Surveying and Mapping Science and Technology, 1991, 16(1): 13-18
    [23]
    Lee V G, Fred J D. Progress in ETA-II magnetic field alignment using stretched wire and low energy electron beam techniques[C]//Proceedings of the Linear Accelerator Conference. 1990, 22(3): 21-29.
    [24]
    Kevan H. Technical proposal for wire position sensor[EB/OL]. http://www.opensourceinstruments.com/WPS/WPS0/
    [25]
    [26]
    Hélène M D. PACMAN project: A new solution for the high-accuracy alignment of accelerator components[C]//Proceedings of the 7th International Particle Accelerator Conference. 2016, 7(3): 31-39.
    [27]
    Wang Wei, Tang Zheng, He Xiaoye, et al. Design and construction of a multi-sensor position monitoring system applied to key components of synchrotron sources[J]. Nucl Sci Tech, 2016, 27: 134. doi: 10.1007/s41365-016-0132-4
    [28]
    Richard S, Erlan B, Lee G. Precision alignment capabilities at LLNL[R]. GEMTN93-339, 1992.
    [29]
    Rodrigo J L, Crhistian R B, Maria L C R, et al. Precision engineering-Magnet alignment on a common girder development of a length[J]. Precision Engineering, 2017(50): 53-62.
    [30]
    Alain L. Principles & status of SOLEIL alignment system[C]//Proceeding of the 8th International Workshops on Accelerator Alignment. 2004, 8(2): 13-19.
    [31]
    Panofsky W K H. The use of a magnetic pickup as an alignment indicator with a stretched wire technique[R]. Tech Note TN-65-74, 1965.
    [32]
    Panofsky W K H. Stretched reference wire magnetic pickup alignment system[P]. US Patent 3, 470, 460, 1969.
    [33]
    Mimashi T, Kuroda S, Nakayama H, et al. The position monitor using stretched wire technique[C]//Proceedings of the 3th International Workshops on Accelerator Alignment. 1993, 3(1): 17-19.
    [34]
    Robert E R. Some alignment considerations for the next linear collider[C]//Proceedings of the 4th International Workshops on Accelerator Alignment. 1995, 4(3): 11-19.
    [35]
    Daniel R. Status of ESRF alignment facilities WPS ready for an automatic 2D smoothing of a storage ring[C]//Proceedings of the 4th International Workshops on Accelerator Alignment. 1995, 4(3): 21-29.
    [36]
    Schwalm. C. Straight line reference system-status report on Poisson system calibration[C]//Proceedings of the 12th International Workshops on Accelerator Alignment. 2012, 12(11): 51-59.
    [37]
    Hélène M D, Bestmann P, Herty A, et al. oWPS versus cWPS[C]//Proceedings of the 12th International Workshops on Accelerator Alignment. 2012, 12(11): 71-79.
    [38]
    Yu C, Karl F, Ilardo M, et al. Smoothing of NSLS-II sotrage ring by girder survey[C]//Proceedings of the 14th International Workshops on Accelerator Alignment. 2016, 14(2): 21-29.
    [39]
    Georg G. Proposal for the alignment of the “Loose End”[C]//Proceedings of the X-Ray FEL Theory and Simulation Codes Workshop. 2006, 1(2): 6-11.
    [40]
    Hyojin C, Sangbong L, Hong-Gi Lee, et al. 2D WPS system for measuring the location changes in real time of PAL-XFEL devices[J]. Journal of the korean physical society, 2018, 73(8): 1034-1041. doi: 10.3938/jkps.73.1034
    [41]
    Hiroaki K, Yuichi O, Morihiro K, et al. Construction and alignment of test half-cell of SPring-8-II[C]//Proceedings of the 15th International Workshops on Accelerator Alignment. 2018, 15(4): 12-26.
    [42]
    Zhang C, Mitsuda C, Kajimoto K. Eigenfrequency wire alignment system for magnet fiducialization[C]//Proceedings of the 14th International Workshops on Accelerator Alignment. 2016, 14(2): 13-26.
    [43]
    Wolf Z, Ruland R, Dix B, et al. Alignment tools used to locate a wire and a laser beam in the VISA[C]//Proceedings of the 6th International Workshops on Accelerator Alignment. 1999, 6(3): 14-25.
    [44]
    Prall M, Johren R, Hannen V, et al. Contactless 2-dim laser sensor for 3-dim wire position and tension measurements[C]//IEEE Nuclear Science Symposium Conference. 2008.
    [45]
    Robert E S. Characteristics of directional coupler beam position monitors[J]. IEEE Transactions on Nuclear Science, 1985, 32(5): 1933-1937. doi: 10.1109/TNS.1985.4333772
    [46]
    Andrey K. Multi-point wide-range precision alignment based on a stretched wire technique[C]//Proceedings of the 3th International Workshops on Accelerator Alignment. 1993, 3(2): 21-29.
    [47]
    De M, Dena J C, Ferianis M, et al. Elettra BPM system (hardware and software): First results[C]//Proceeding of the European Particle Accelerator Conference. 1994.
    [48]
    Eddy N, Fellenz B, Prieto P, et al. A wire position monitor system for the 1.3 GHz Tesla-style cryomodule at the Fermilab new-muon-lab accelerator[C]//15th International Conference on RF Superconductivity(SRF2011). 2011: 25-29.
    [49]
    Rawnsley W R, Bylinski Y, Fong K, et al. Alignment of the ISAC-II medium beta cryostat with a wire monitoring system[C]//The Proceedings of the Twentieth International Cryogenic Engineering Conference(ICEC20). 2005: 931-934.
    [50]
    Bertolini A. Vibration diagnostics instrumentation for ILC[J]. Measurement Science & Technology, 2007, 8(18): 2293-2298.
    [51]
    Zhang D, Eddy N, Fellenz B, et al. A wire position monitor system for superconducting cryostat at Fermilab[C]//The 2010 Beam Instrumentation Workshop. 2010, 2(3): 2-6.
    [52]
    Hitoshi H. Superconducting accelerator development at STF for ILC[C]//Proceedings of the 12th Annual Meeting of Particle Accelerator Society of Japan. 2015.
    [53]
    Zhu H, Dong L, Li B. Design and simulation of a wire position monitor for cryogenic systems in an ADS linac[J]. Chinese Physics C, 2014, 38(3): 808-901.
    [54]
    Xu T, Ao H, Bird B, et al. FRIB cryomodule design and production[C]//Proceedings of LINAC, East Lansing. 2016.
    [55]
    Robert M M. Alignment of drift tube quadrupole lenses using a pulsed wire[C]//Proceedings of the 4th International Conference on Magnet Technology. 1972.
    [56]
    Warren R W. Limitations on the use of the pulsed-wire field measuring technique[J]. Nuclear Instruments and Methods in Physics Research A, 1988, 23(272): 257-263.
    [57]
    Dino Z, Richard P, Walker A. A stretched wire system for accurate integrated magnetic field measurements in insertion devices[J]. Nuclear Instruments and Methods in Physics Research A, 1996, 24(376): 275-282.
    [58]
    Alexander T. Vibrating wire field-measuring technique[J]. Nuclear Instruments and Methods in Physics Research A, 1997, 43(399): 185-194.
    [59]
    Arutunian S G, Bakshetyan K G, Dobrovolsky N M, et al. Petra proton beam profiling by vibrating wire scanner[C]//Proceedings of DIPAC.2005.
    [60]
    Arutunian S G, Davtyan M A, Vasiniuk I E, et al. Impact of environmental variability on vibrating wire monitor operation[C]//Proceedings of DIPAC09. 2009.
    [61]
    Jain A, Anerella M, Ganetis G, et al. Vibrating wire R&D for alignment of multipole magnets in NSLS-II[C]//Proceedings of the 10th International Workshop on Accelerator Alignment. 2008.
    [62]
    Alexander T, Yurii L, Wolf Z. A study of undulator magnets characterization using the vibrating wire technique[J]. Nuclear Instruments and Methods in Physics Research Section A, 2010, 622(3): 650-656. doi: 10.1016/j.nima.2010.06.362
    [63]
    Wouters C. Vibrating wire technique and phase lock loop for finding the magnetic axis of quadrupoles[J]. IEEE Transactions on Applied Superconductivity: A Publication of the IEEE Superconductivity Committee, 2012, 3(3): 1-4.
    [64]
    Bec G L, Chavanne J, Penel C. Stretched wire measurement of multipole accelerator magnets[J]. Physical Review Special Topics—Accelerators and Beams, 2012, 15: 022401. doi: 10.1103/PhysRevSTAB.15.022401
    [65]
    Caughan M M, Tiefenback M, Turner D. Improvements to existing Jefferson LAB wire scanners[C]//Proceedings of IPAC.2013.
    [66]
    Arutunian S G, Avetisyan A E, Davtyan M M, et al. Large aperture vibrating wire monitor with two mechanically coupled wires for beam halo measurements[J]. Physical Review Special Topics—Accelerators and Beams, 2014, 17: 032802. doi: 10.1103/PhysRevSTAB.17.032802
    [67]
    Caiazza D, Lasheras N C, Hélène M D, et al. New solution for the high accuracy alignment of accelerator components[J]. Physical Review Accelerators and Beams, 2017, 20: 083501. doi: 10.1103/PhysRevAccelBeams.20.083501
    [68]
    吴蕾. 振动线测量技术研究[D]. 北京: 中国科学院大学, 2016.

    Wu Lei. Research of vibrating wire measurement technique[D]. Beijing: The University of Chinese Academy of Sciences, 2016
    [69]
    Mezentsev N, Tsukanov V, Zorin A. Magnetic measurements of superconducting insertion devices by stretched wire with direct current[C]//International Conference on Synchrotron and Free Electron Laser Radiation: Generation and Application. 2016: 4-8.
    [70]
    Tsechuan T. Vibrating wire method and related positioning study at NSRRC[C]//Proceedings of the 15th International Workshops on Accelerator Alignment. 2018.
    [71]
    Vinit K, Ganeshwar M. Analysis of pulsed wire method for field integral measurements in undulators[J]. PRAMANA Journal of Physics, 2010, 74(5): 743-753. doi: 10.1007/s12043-010-0095-7
    [72]
    Renan G, Rodrigo J L, Geraldo C, et al. New sensor and non-contact geometrical survey for the vibrating wire technique[J]. Nuclear Instruments and Methods in Physics Research A, 2016, 811: 115-123. doi: 10.1016/j.nima.2015.12.016
    [73]
    Coosemans W, Dehning B, Hildreth M, et al. Performance of wire position sensors in a radiation environment[C]//Proceedings of the 6th International Workshops on Accelerator Alignment.1999.
    [74]
    Prochnow J, Coosemans W, Dehning B, et al. Determination of the accuracy of wire position sensor[C]//Proceedings of EPAC. 2000.
    [75]
    Rodrigo J L, Crhistian R B, Maria L C, et al. Engineering survey planning for the alignment of a particle accelerator: part I. Proposition of an assessment method[J]. Measurement Science and Technology, 2018, 29: 034006. doi: 10.1088/1361-6501/aa9dc4
    [76]
    Friedsam H, Oren W, Pietryka M. Magnet fiducialization with coordinate measuring machines[C]//Proceedings of the 1th International Workshops on Accelerator Alignment. 1989, 21(2): 22-34.
    [77]
    Claude S. A rotary mount for submicrometric positioning of a stretched wire axis within a coordinate measuring machine[C]//Proceedings of the 14th International Workshops on Accelerator Alignment. 2016.
    [78]
    Dirk M, Dominique M, Camille V. Recent developments for a photogrammetric system to measure offsets w.r.t. stretched wires[C]//Proceedings of the 15th International Workshops on Accelerator Alignment. 2018.
    [79]
    Pasquale A, Domenico C, Carlo P, et al. Performance of the stretched- and vibrating-wire techniques and correction of background fields in locating quadruple magnetic axes[C]//XXI IMEKO World Congress “Measurement in Research and Industry”. 2015.
    [80]
    Arpaia P, Artoos K, Buzio M, et al. Sensors applications within the research framework of the PACMAN Project on metrology for particle accelerators at CERN[C]//IEEE 2016 Sensors Application Symposium. 2016.
    [81]
    Nosych A A, Iriso U, Olmos A, et al. Measurements and calibration of the stripline bpm for the ELI-NP facility with the stretched wire method[C]//Proceedings of IBIC. 2015.
    [82]
    Durand S, Touzé T, Rude V, et al. Improving the geometry of a 3d longitudinal network using a stretched wire[C]//Proceedings of the 15th International Workshops on Accelerator Alignment. 2018: 34-39
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Figures(9)

    Article views (1804) PDF downloads(74) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return