The diagnosis of X-ray flux with high temporal and spatial resolutions is important in the researches of hohlraum physics, and spectral response coupling and design of spectrum rebuilding unit are key issues in the design of flat-response streak camera. Based on the study of the spectral response characteristics of streak camera and X-ray transmission photocathode, compound filter is adopted to realize the M-band band-pass flat-response of streak camera. The calculation shows that the flatness of spectral response is better than 2.7% with an acceptable band pass effect.