wei min-xi, yang jia-min, yang guo-hong. Calibration of integral diffraction coefficient of flat crystal[J]. High Power Laser and Particle Beams, 2006, 18.
Citation:
wei min-xi, yang jia-min, yang guo-hong. Calibration of integral diffraction coefficient of flat crystal[J]. High Power Laser and Particle Beams, 2006, 18.
wei min-xi, yang jia-min, yang guo-hong. Calibration of integral diffraction coefficient of flat crystal[J]. High Power Laser and Particle Beams, 2006, 18.
Citation:
wei min-xi, yang jia-min, yang guo-hong. Calibration of integral diffraction coefficient of flat crystal[J]. High Power Laser and Particle Beams, 2006, 18.
The flat crystal spectrometer is one of the fundamental tools to diagnose X-ray. Quasi-quantitative and quantitative analyses of X-ray spectra require calibrations of the flat crystal’s characteristic parameters, such as integral reflective coefficient and FWHM of the flat crystal’s rocking curve. Using BSRF 3B3 beam as the X-ray source, this article gives an indirect method to calibrate the flat PET crystal. Good results are obtained in the range of 2.1~6.0 keV and interrelated analysis is provided.