Volume 15 Issue 07
Jul.  2003
Turn off MathJax
Article Contents
hu meng chun, ye weng ying, zhou dian zhong, et al. DD neutron relative sensitivity of cerium fluoride scintillation detector[J]. High Power Laser and Particle Beams, 2003, 15.
Citation: hu meng chun, ye weng ying, zhou dian zhong, et al. DD neutron relative sensitivity of cerium fluoride scintillation detector[J]. High Power Laser and Particle Beams, 2003, 15.

DD neutron relative sensitivity of cerium fluoride scintillation detector

  • Publish Date: 2003-07-15
  • Two kinds scintillation detectors are composed of cerium fluoride (CeF3) and ST401 with photomultiplier tube respectively, the experimental sample of CeF3 is the newly developed inorganic scintillator in China in recent years, ST401 is a common plastic organic scintillator. The DD neutron relative sensitivity of the two scintillation detectors are measured in DD neutrons source radiation field, which indicate that the DD neutron sensitivity of CeF3 scintillation detector is less than onetenth that of ST401 with same volume.
  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索
    Article views (2147) PDF downloads(637) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return