Volume 17 Issue 03
Mar.  2005
Turn off MathJax
Article Contents
song zhao-hui, ruan xi-chao, dai qiu-sheng, et al. Principle and application of sensitivity calibration by single particle for current detector[J]. High Power Laser and Particle Beams, 2005, 17.
Citation: song zhao-hui, ruan xi-chao, dai qiu-sheng, et al. Principle and application of sensitivity calibration by single particle for current detector[J]. High Power Laser and Particle Beams, 2005, 17.

Principle and application of sensitivity calibration by single particle for current detector

  • Publish Date: 2005-03-15
  • A method adopting the charge analog digital conversion technique to calibrate a current scintillation detector is described. Each pulse charge produced by a single particle is collected and recorded. Calculating the mean current, the sensitivity of the detector is measured. The results agree with the measurement presented by the traditional direct current method within ±8%. This new method can get higher accuracy, subtract background easily and it does not depend on the intensity of the radiation source. By applying time-of-flight method, it is capable of particle discrimination. It can separate the pulse charge produced by neutron from that produced by gamma in a neutron-gamma mixed field. The sensitivity of current detector to each type of particle can be measured separately. This metho
  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索
    Article views (2207) PDF downloads(544) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return